Decision
Substantially Equivalent
Decision date
Feb 26, 2026
Received
Jan 27, 2026
510(k) type
Traditional
Product code
FLESterilizer, Steam
Review panel
General Hospital
Device class
Class II
Regulation
21 CFR 880.6880
Third-party review
Yes
Summary or statement
Summary
Review time
30 daysreceipt to decision, calendar days
Median in FLE
140 daysthe latest 24 other decisions
FLE decisions
5in the last 12 months
Applicant’s 510(k)s
3in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Enbio PRO is a table-top steam sterilizer, a benchtop machine that cleans medical and dental instruments with pressurized steam. It uses an air-removal (pre-vacuum) process, meaning it pumps air out of the sealed chamber before steam is added so the steam can reach into the instruments. The device has a heated aluminum chamber, a perforated tray for the load, and a touch screen for choosing a program.
What it’s cleared for
It is cleared for use by health care providers to sterilize dental and medical instruments that are validated to be sterilized by steam. It is not designed for liquid loads, biomedical waste, or materials that cannot tolerate steam sterilization. The summary lists three programs: 134°C for 4 minutes, 121°C for 30 minutes, and a 134°C FAST immediate-use cycle for 4 minutes, each with a maximum load of 1.6 kg (3.52 lbs).
Compared with (the predicate)
How it differs
The Enbio PRO has a larger sterilization chamber (5.3 L versus 2.7 L) and a higher maximum load (1.6 kg versus 0.5 kg). It is also heavier (about 20 kg versus 15 kg) and has somewhat different exposure and drying times, which the company says fall within standard values for each temperature. Its firmware was updated to support these differences.
Testing the summary lists
  • Electrical safety (IEC 61010-1, IEC 61010-2-040)
  • Electromagnetic compatibility (IEC 60601-1-2)
  • Air removal: vacuum test and Bowie-Dick test (AAMI ANSI ST55:2016)
  • Temperature and pressure: full cycle and half cycle studies (AAMI ANSI ST55:2016)
  • Biological indicator challenges, including dental turbine testing (AAMI ANSI ST55:2016)
  • Firmware validation using the same method used for the predicate device

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K260254 took 30 calendar days from receipt to decision. Of the latest 24 other decisions in product code FLE, 22 took longer, 2 were faster; their median was 140 days.

Latest decisions in FLE

NumberDevice and applicantDecision and review time
K260354 Foster Plus-22L (SNK/CE-F22)Zhejiang Fomos Medical Technology Co., Ltd. · Traditional Oct 30, 2026269 days
K261055 AMSCO 700 Steam SterilizerSTERIS Corporation · Traditional Jun 24, 202685 days
K260254 Enbio PROEnbio Group AG · Traditional Feb 26, 202630 days
K252307 Getinge GSS67N Series Steam SterilizerMaquet GmbH (A Getinge Group Company) · Traditional Jan 22, 2026182 days
K250168 Life Ultra 25L Tabletop Mono Chamber Steam SterilizerLucas Lifecare · Traditional Dec 30, 2025343 days
K250164 Cassette Autoclave (ACA5)Guangzhou Ajax Medical Equipment Co., Ltd. · Traditional Aug 14, 2025205 days
K243994 Steam Sterilizer (2545D)Ningbo Ican Machines Co., Ltd. · Traditional Jun 6, 2025162 days
K243801 FRONT-LINE Field Sterilizer (FL135)Fort Defiance Industries, LLC · Special Mar 7, 202586 days
K233026 Midmark Smart M9® Sterilizer, Midmark Smart M11® SterilizerMidmark Corporation · Traditional Apr 4, 2024195 days
K233504 Lexa MINI (RIS-303, RIS-305)W&H Sterilization S.R.L. · Traditional Feb 8, 2024100 days

Other 510(k)s from Enbio Group AG

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K213991 Enbio SEnbio Group AG · Special Feb 22, 202263 days
K210279 Enbio SEnbio Group AG · Traditional Oct 29, 2021270 days

Watch FLE in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get FLE every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in FLE. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.