FDA device news · Product code FLE
FDA 510(k) clearance
Enbio PRO (K260254): Enbio Group AG 510(k) clearance
Oensingen, CH
- Decision
- Substantially Equivalent
- Decision date
- Feb 26, 2026
- Received
- Jan 27, 2026
- 510(k) type
- Traditional
- Product code
- FLESterilizer, Steam
- Review panel
- General Hospital
- Device class
- Class II
- Regulation
- 21 CFR 880.6880
- Third-party review
- Yes
- Summary or statement
- Summary
- Review time
- 30 daysreceipt to decision, calendar days
- Median in FLE
- 140 daysthe latest 24 other decisions
- FLE decisions
- 5in the last 12 months
- Applicant’s 510(k)s
- 3in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Enbio PRO is a table-top steam sterilizer, a benchtop machine that cleans medical and dental instruments with pressurized steam. It uses an air-removal (pre-vacuum) process, meaning it pumps air out of the sealed chamber before steam is added so the steam can reach into the instruments. The device has a heated aluminum chamber, a perforated tray for the load, and a touch screen for choosing a program.
- What it’s cleared for
- It is cleared for use by health care providers to sterilize dental and medical instruments that are validated to be sterilized by steam. It is not designed for liquid loads, biomedical waste, or materials that cannot tolerate steam sterilization. The summary lists three programs: 134°C for 4 minutes, 121°C for 30 minutes, and a 134°C FAST immediate-use cycle for 4 minutes, each with a maximum load of 1.6 kg (3.52 lbs).
- Compared with (the predicate)
- K213991, Enbio S
- How it differs
- The Enbio PRO has a larger sterilization chamber (5.3 L versus 2.7 L) and a higher maximum load (1.6 kg versus 0.5 kg). It is also heavier (about 20 kg versus 15 kg) and has somewhat different exposure and drying times, which the company says fall within standard values for each temperature. Its firmware was updated to support these differences.
- Testing the summary lists
- Electrical safety (IEC 61010-1, IEC 61010-2-040)
- Electromagnetic compatibility (IEC 60601-1-2)
- Air removal: vacuum test and Bowie-Dick test (AAMI ANSI ST55:2016)
- Temperature and pressure: full cycle and half cycle studies (AAMI ANSI ST55:2016)
- Biological indicator challenges, including dental turbine testing (AAMI ANSI ST55:2016)
- Firmware validation using the same method used for the predicate device
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K260254 took 30 calendar days from receipt to decision. Of the latest 24 other decisions in product code FLE, 22 took longer, 2 were faster; their median was 140 days.
Latest decisions in FLE
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260354 | Foster Plus-22L (SNK/CE-F22)Zhejiang Fomos Medical Technology Co., Ltd. · Traditional | Oct 30, 2026269 days |
| K261055 | AMSCO 700 Steam SterilizerSTERIS Corporation · Traditional | Jun 24, 202685 days |
| K260254 | Enbio PROEnbio Group AG · Traditional | Feb 26, 202630 days |
| K252307 | Getinge GSS67N Series Steam SterilizerMaquet GmbH (A Getinge Group Company) · Traditional | Jan 22, 2026182 days |
| K250168 | Life Ultra 25L Tabletop Mono Chamber Steam SterilizerLucas Lifecare · Traditional | Dec 30, 2025343 days |
| K250164 | Cassette Autoclave (ACA5)Guangzhou Ajax Medical Equipment Co., Ltd. · Traditional | Aug 14, 2025205 days |
| K243994 | Steam Sterilizer (2545D)Ningbo Ican Machines Co., Ltd. · Traditional | Jun 6, 2025162 days |
| K243801 | FRONT-LINE Field Sterilizer (FL135)Fort Defiance Industries, LLC · Special | Mar 7, 202586 days |
| K233026 | Midmark Smart M9® Sterilizer, Midmark Smart M11® SterilizerMidmark Corporation · Traditional | Apr 4, 2024195 days |
| K233504 | Lexa MINI (RIS-303, RIS-305)W&H Sterilization S.R.L. · Traditional | Feb 8, 2024100 days |
Other 510(k)s from Enbio Group AG
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
Watch FLE in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get FLE every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in FLE. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K260254Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:FLERetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:FLERetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:FLE AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Enbio Group AG"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.