FDA device news · Product code PPN
FDA 510(k) clearance
SEISMIQ™ Intravascular Lithotripsy System (K260246): Bolt Medical, Inc. 510(k) clearance
Carlsbad, CA
- Decision
- Substantially Equivalent
- Decision date
- Aug 24, 2026
- Received
- Jan 27, 2026
- 510(k) type
- Traditional
- Product code
- PPNPercutaneous Catheter, Ultrasound
- Review panel
- Cardiovascular
- Device class
- Class II
- Regulation
- 21 CFR 870.1250
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 209 daysreceipt to decision, calendar days
- Median in PPN
- 56 daysthe latest 13 other decisions
- PPN decisions
- 2in the last 12 months
- Applicant’s 510(k)s
- 2in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The SEISMIQ Intravascular Lithotripsy System is a balloon catheter paired with a reusable console. The catheter is delivered through the arteries of the leg to the lesion, and emitters inside the balloon generate pulsed mechanical energy (lithotripsy) that breaks up calcium in the blockage before the balloon is opened at low pressure.
- What it’s cleared for
- The system is cleared for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral arteries of the leg, including the iliac, femoral, iliofemoral, popliteal, and infra-popliteal arteries. It is not for use in the coronary, carotid, or cerebral arteries.
- Compared with (the predicate)
- K250225, SEISMIQ Intravascular Lithotripsy System (previously named the Bolt IVL System)
- How it differs
- The new catheters are added to the existing line in smaller balloon diameters and lengths to widen the size range. The summary states that the intended use, principles of operation, energy source, shelf life, and sterilization are the same as the predicate.
- Testing the summary lists
- Bench testing of catheter dimensions, strength, flexibility, balloon performance, and fatigue
- Fluoroscopic visibility and particulate evaluation
- Pulsing cycles and sonic output
- Biocompatibility (ISO 10993-1:2020, leveraged from predicate K250225)
- Animal testing (chronic GLP porcine study, leveraged from K250225)
- Clinical data (RESTORE ATK and RESTORE BTK studies, leveraged from K250225)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K260246 took 209 calendar days from receipt to decision. Of the latest 13 other decisions in product code PPN, 13 were faster; their median was 56 days.
Latest decisions in PPN
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260246 | SEISMIQ™ Intravascular Lithotripsy SystemBolt Medical, Inc. · Traditional | Aug 24, 2026209 days |
| K262201 | Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL CatheterShockwave Medical, Inc. · Special | Jul 27, 202628 days |
| K250225 | Bolt Intravascular Lithotripsy (IVL) SystemBolt Medical, Inc. · Traditional | Mar 25, 202557 days |
| K242213 | Shockwave Intravascular Lithotripsy (IVL) System with the Javelin Peripheral Intravascular Lithotripsy (IVL) CatheterShockwave Medical, Inc. · Traditional | Sep 27, 202460 days |
| K240954 | Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave L6 Peripheral IVL CatheterShockwave Medical, Inc. · Special | May 7, 202429 days |
| K240225 | Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL CatheterShockwave Medical, Inc. · Traditional | Mar 22, 202456 days |
| K221852 | Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave L6 Peripheral IVL CatheterShockwave Medical, Inc. · Traditional | Aug 16, 202250 days |
| K221041 | Shockwave Medical Peripheral Intravascular Lithotripsy (IVL) SystemShockwave Medical, Inc. · Special | Apr 25, 202217 days |
| K203365 | Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave M5+ Peripheral IVL CatheterShockwave Medical, Inc. · Traditional | Apr 22, 2021157 days |
| K191840 | Peripheral Intravascular Lithotripsy (IVL) Catheters, IVL Generator and Connector Cable Kit, IVL Connector Cable (Replacement)Shockwave Medical, Inc. · Special | Aug 7, 201929 days |
Other 510(k)s from Bolt Medical, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K250225 | Bolt Intravascular Lithotripsy (IVL) SystemBolt Medical, Inc. · Traditional | Mar 25, 202557 days |
Watch PPN in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get PPN every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in PPN. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K260246Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:PPNRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:PPNRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:PPN AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Bolt Medical, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil