FDA device news · Product code PPN

FDA 510(k) clearance

SEISMIQ™ Intravascular Lithotripsy System (K260246): Bolt Medical, Inc. 510(k) clearance

Carlsbad, CA

Follow PPN: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Aug 24, 2026
Received
Jan 27, 2026
510(k) type
Traditional
Product code
PPNPercutaneous Catheter, Ultrasound
Review panel
Cardiovascular
Device class
Class II
Regulation
21 CFR 870.1250
Third-party review
No
Summary or statement
Summary
Review time
209 daysreceipt to decision, calendar days
Median in PPN
56 daysthe latest 13 other decisions
PPN decisions
2in the last 12 months
Applicant’s 510(k)s
2in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The SEISMIQ Intravascular Lithotripsy System is a balloon catheter paired with a reusable console. The catheter is delivered through the arteries of the leg to the lesion, and emitters inside the balloon generate pulsed mechanical energy (lithotripsy) that breaks up calcium in the blockage before the balloon is opened at low pressure.
What it’s cleared for
The system is cleared for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral arteries of the leg, including the iliac, femoral, iliofemoral, popliteal, and infra-popliteal arteries. It is not for use in the coronary, carotid, or cerebral arteries.
Compared with (the predicate)
  • K250225, SEISMIQ Intravascular Lithotripsy System (previously named the Bolt IVL System)
How it differs
The new catheters are added to the existing line in smaller balloon diameters and lengths to widen the size range. The summary states that the intended use, principles of operation, energy source, shelf life, and sterilization are the same as the predicate.
Testing the summary lists
  • Bench testing of catheter dimensions, strength, flexibility, balloon performance, and fatigue
  • Fluoroscopic visibility and particulate evaluation
  • Pulsing cycles and sonic output
  • Biocompatibility (ISO 10993-1:2020, leveraged from predicate K250225)
  • Animal testing (chronic GLP porcine study, leveraged from K250225)
  • Clinical data (RESTORE ATK and RESTORE BTK studies, leveraged from K250225)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K260246 took 209 calendar days from receipt to decision. Of the latest 13 other decisions in product code PPN, 13 were faster; their median was 56 days.

Latest decisions in PPN

NumberDevice and applicantDecision and review time
K260246 SEISMIQ™ Intravascular Lithotripsy SystemBolt Medical, Inc. · Traditional Aug 24, 2026209 days
K262201 Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL CatheterShockwave Medical, Inc. · Special Jul 27, 202628 days
K250225 Bolt Intravascular Lithotripsy (IVL) SystemBolt Medical, Inc. · Traditional Mar 25, 202557 days
K242213 Shockwave Intravascular Lithotripsy (IVL) System with the Javelin Peripheral Intravascular Lithotripsy (IVL) CatheterShockwave Medical, Inc. · Traditional Sep 27, 202460 days
K240954 Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave L6 Peripheral IVL CatheterShockwave Medical, Inc. · Special May 7, 202429 days
K240225 Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave E8 Peripheral IVL CatheterShockwave Medical, Inc. · Traditional Mar 22, 202456 days
K221852 Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave L6 Peripheral IVL CatheterShockwave Medical, Inc. · Traditional Aug 16, 202250 days
K221041 Shockwave Medical Peripheral Intravascular Lithotripsy (IVL) SystemShockwave Medical, Inc. · Special Apr 25, 202217 days
K203365 Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave M5+ Peripheral IVL CatheterShockwave Medical, Inc. · Traditional Apr 22, 2021157 days
K191840 Peripheral Intravascular Lithotripsy (IVL) Catheters, IVL Generator and Connector Cable Kit, IVL Connector Cable (Replacement)Shockwave Medical, Inc. · Special Aug 7, 201929 days

Other 510(k)s from Bolt Medical, Inc.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K250225 Bolt Intravascular Lithotripsy (IVL) SystemBolt Medical, Inc. · Traditional Mar 25, 202557 days

Watch PPN in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get PPN every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in PPN. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.