FDA device news · Product code MUJ
FDA 510(k) clearance
Multimodality Simulation Workspace (MM Sim) (v1.0.0) (K260207): Philips Medical Systems Nederland B.V. 510(k) clearance
Best, NL
- Decision
- Substantially Equivalent
- Decision date
- May 27, 2026
- Received
- Jan 23, 2026
- 510(k) type
- Traditional
- Product code
- MUJSystem, Planning, Radiation Therapy Treatment
- Review panel
- Radiology
- Device class
- Class II
- Regulation
- 21 CFR 892.5050
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 124 daysreceipt to decision, calendar days
- Median in MUJ
- 164 daysthe latest 24 other decisions
- MUJ decisions
- 12in the last 12 months
- Applicant’s 510(k)s
- 108in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- Multimodality Simulation Workspace (MM Sim) is a software-only medical device that supports radiation therapy simulation and preparation. It provides 3D and 4D image viewing, combines images from different imaging types (multimodality image manipulation), and offers tools to draw outlines of body regions (contouring) using images that were already acquired.
- What it’s cleared for
- MM Sim is cleared to help prepare radiation treatments, such as brachytherapy, particle therapy, and external beam radiation therapy. It lets users view and edit contours of body regions such as skin, targets, and organs-at-risk, and create simple geometric treatment plans. Contours, images, and treatment plans can be exported to a Treatment Planning System.
- Compared with (the predicate)
- K232799, syngo.RT Image Suite
- How it differs
- The summary says MM Sim uses atlas-based and model-based assistive segmentation, where the predicate is described as using deep learning-based auto-segmentation. MM Sim also adds 4D CT support for mid-ventilation and mid-position datasets and uses configurable navigation pathways, where the predicate uses predefined templates.
- Testing the summary lists
- Software verification and validation, including requirements-based verification and simulated-use validation (IEC 62304, ISO 14971)
- Image data handling and interoperability, including DICOM import, processing, and export
- Visualization, registration, manual contouring, DICOM RT import/export, and end-to-end workflow execution
- Image quality using phantom-based and representative datasets (spatial resolution, noise, uniformity, artifacts, extended field-of-view)
- Mid-Position and Mid-Ventilation dataset generation, with independent expert review of Mid-P image quality
- Navigation pathway functionality and atlas-based and model-based assistive segmentation
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K260207 took 124 calendar days from receipt to decision. Of the latest 24 other decisions in product code MUJ, 14 took longer, 8 were faster, 2 took the same time; their median was 164 days.
Latest decisions in MUJ
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261655 | RT Elements (4.6)Brainlab SE · Traditional | Oct 5, 2026139 days |
| K260750 | Virtual Cone (1.0)Varian Medical Systems, Inc. · Traditional | Jul 8, 2026124 days |
| K253050 | RadOncAIInformai · Traditional | Jun 18, 2026269 days |
| K260207 | Multimodality Simulation Workspace (MM Sim) (v1.0.0)Philips Medical Systems Nederland B.V. · Traditional | May 27, 2026124 days |
| K252784 | eXaSkin PlusAnatomical Geometry S.L. · Traditional | May 22, 2026262 days |
| K253078 | ART.1-USSeetreat Pty, Ltd. · Traditional | May 14, 2026233 days |
| K252304 | syngo.via RT Image Suite VC10Varian Medical Systems, Inc. · Traditional | Mar 18, 2026237 days |
| K252002 | Monaco RTP System (6.3)Elekta Solutions AB · Traditional | Feb 19, 2026237 days |
| K260308 | TrueFit BolusAdaptiiv Medical Technologies, Inc. · Special | Feb 19, 202620 days |
| K253091 | ART-Plan+ (v3.1.0)Therapanacea Sas · Traditional | Dec 23, 202591 days |
Other 510(k)s from Philips Medical Systems Nederland B.V.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K254205 | Zenition 30Philips Medical Systems Nederland B.V. · Traditional | Aug 7, 2026221 days |
| K262330 | dS Breast 7ch 1.5TdS Breast 7ch 3.0TdS Breast 16ch 1.5TdS Breast 16ch 3.0TdS Breast Coil 7ch 1.5TdS Sentinelle Breast 16ch 1.5TdS Sentinelle Breast 16ch 3.0TPhilips Medical Systems Nederland B.V. · Special | Jul 30, 202621 days |
| K262294 | dS FootAnkle 16Ch 1.5T; dS FootAnkle 16Ch 3.0T; dS HiRes Hand/Wrist 16Ch 1.5T; dS HiRes Hand/Wrist 16CH 3.0T; dS Small Extremity 16Ch 1.5T; dS Small Extremity 16Ch 3.0TPhilips Medical Systems Nederland B.V. · Special | Jul 24, 202617 days |
| K253614 | EchoNavigator R5.0Philips Medical Systems Nederland B.V. · Traditional | Mar 17, 2026119 days |
| K260169 | AV Cardiac CTPhilips Medical Systems Nederland B.V. · Traditional | Mar 5, 202644 days |
| K253648 | Ingenia, Ingenia CX, Ingenia Elition S/X, Ingenia Ambition S/X, BlueSeal SE/XE/QE, MR 5300 and MR 7700 MR Systems including UpgradesPhilips Medical Systems Nederland B.V. · Abbreviated | Feb 23, 202695 days |
| K253735 | AV VascularPhilips Medical Systems Nederland B.V. · Traditional | Jan 22, 202659 days |
| K254190 | dS Base 1.5T; dS Base 3.0T; dS Head 1.5T; dS Head 3.0T; dS HeadNeck 1.5T; dS HeadNeck 3.0TPhilips Medical Systems Nederland B.V. · Special | Jan 14, 202622 days |
| K252645 | LumiGuide Equipment R2.1; LumiGuide Wire; LumiGuide 3D HubPhilips Medical Systems Nederland B.V. · Abbreviated | Oct 24, 202564 days |
| K251215 | Philips IntelliSpace CardiovascularPhilips Medical Systems Nederland B.V. · Traditional | Oct 2, 2025167 days |
| K250181 | AV ViewerPhilips Medical Systems Nederland B.V. · Traditional | Jul 15, 2025174 days |
| K251808 | Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX; Ingenia 3.0T CX; Ingenia Elition S; Ingenia Elition X; Ingenia Ambition S; Ingenia Ambition X; and MR 5300 MR SystemsPhilips Medical Systems Nederland B.V. · Special | Jul 11, 202529 days |
Watch MUJ in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get MUJ every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in MUJ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K260207Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:MUJRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:MUJRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:MUJ AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Philips Medical Systems Nederland B.V."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.