FDA device news · Product code DQX
FDA 510(k) clearance
Small Diameter Core Wire (CMAN-000-00, CMAN-000-14) (K260168): Skycade Med Corporation 510(k) clearance
Dallas, TX
- Decision
- Substantially Equivalent
- Decision date
- Oct 16, 2026
- Received
- Jan 20, 2026
- 510(k) type
- Traditional
- Product code
- DQXWire, Guide, Catheter
- Review panel
- Cardiovascular
- Device class
- Class II
- Regulation
- 21 CFR 870.1330
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 269 daysreceipt to decision, calendar days
- Median in DQX
- 156 daysthe latest 24 other decisions
- DQX decisions
- 12in the last 12 months
- Applicant’s 510(k)s
- 1in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Small Diameter Core Wire is a guide wire, a thin flexible wire that is threaded into blood vessels to create a path for catheters and other interventional tools. It has a nitinol core wire (a nickel-titanium alloy) with a radiopaque coil at the tip, which shows up on medical imaging so the physician can track its position. It is supplied sterile for single use and is sterilized with ethylene oxide gas.
- What it’s cleared for
- It is cleared for use in percutaneous procedures (procedures done through the skin) to introduce and position catheters and other interventional devices within the peripheral vasculature, meaning the blood vessels outside the heart and brain. The summary says it is intended for use by healthcare professionals in hospitals.
- Compared with (the predicate)
- K024021, Nitrex TM Nitinol Guidewire
- How it differs
- The summary says the device is available in 40, 50 and 80 cm lengths and two diameters (0.014 and 0.018 inch), while the predicate comes in more lengths and diameters and is also indicated for coronary vasculature. The coil is made of stainless steel and gold instead of nitinol, which the summary lists as the one material difference.
- Testing the summary lists
- Sterilization residuals (ISO 10993-7) and endotoxin levels
- Accelerated shelf-life testing (equivalent to four years of aging)
- Animal testing: none performed
- Clinical testing: none performed, because the summary says indications and fundamental technology did not change
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K260168 took 269 calendar days from receipt to decision. Of the latest 24 other decisions in product code DQX, 1 took longer, 22 were faster, 1 took the same time; their median was 156 days.
Latest decisions in DQX
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260168 | Small Diameter Core Wire (CMAN-000-00, CMAN-000-14)Skycade Med Corporation · Traditional | Oct 16, 2026269 days |
| K253146 | Hyperdrive 35 GuidewireRapid-Medical , Ltd. · Traditional | Sep 18, 2026358 days |
| K261281 | CROSSLEAD Standard 035 and CROSSLEAD Stiff 035Filmecc Co., Ltd. · Traditional | Sep 3, 2026139 days |
| K253667 | Accoat Guide WireSp Medical A/S · Traditional | Jul 14, 2026235 days |
| K261640 | Merit MAK Nitinol/Gold-Tungsten Guide WireMerit Medical Ireland , Ltd. · Special | Jul 13, 202656 days |
| K254137 | InQwire Amplatz Guide WireMerit Medical Ireland, Ltd. · Traditional | May 22, 2026151 days |
| K260544 | FMD Peripheral Guide Wire F-14 Flex 6; FMD Peripheral Guide Wire F-14 Tapered 40; FMD Peripheral Guide Wire F-18 Flex SP; FMD Peripheral Guide Wire F-18 DP 25; FMD Peripheral Guide Wire F-18 Tapered 40FMD Co., Ltd. · Special | Mar 20, 202630 days |
| K253746 | Enroute 0.014'' Transcarotid GuidewireLake Region Medical · Special | Mar 19, 2026114 days |
| K253262 | EmeryGlide™ (EG18008901)Nano4imaging GmbH · Special | Mar 6, 2026158 days |
| K253847 | Splashwire Hydrophilic Guide Wire (MSWSTD35150J3)Merit Medical Ireland, Ltd. · Special | Jan 31, 202660 days |
Other 510(k)s from Skycade Med Corporation
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
This is the only 510(k) in openFDA under this applicant name.
Watch DQX in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get DQX every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in DQX. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K260168Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:DQXRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:DQXRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:DQX AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Skycade Med Corporation"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil