FDA device news · Product code JAK
FDA 510(k) clearance
Photonova Spectra, Photonova Spectra Select (K253520): Ge Medical Systems, LLC 510(k) clearance
Waukesha, WI
- Decision
- Substantially Equivalent
- Decision date
- Mar 20, 2026
- Received
- Nov 12, 2025
- 510(k) type
- Traditional
- Product code
- JAKSystem, X-Ray, Tomography, Computed
- Review panel
- Radiology
- Device class
- Class II
- Regulation
- 21 CFR 892.1750
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 128 daysreceipt to decision, calendar days
- Median in JAK
- 168 daysthe latest 24 other decisions
- JAK decisions
- 21in the last 12 months
- Applicant’s 510(k)s
- 113in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- Photonova Spectra and Photonova Spectra Select are X-ray computed tomography (CT) scanners that create cross-sectional images of the body by computer reconstruction of X-ray transmission data. They use a new silicon-based photon counting detector that directly converts X-ray photons into electrical signals and sorts them into 8 energy bins, so each scan captures multi-energy data.
- What it’s cleared for
- The system is cleared to produce cross-sectional body images for head, whole body, cardiac, and vascular CT applications, for patients of all ages. It natively generates high resolution monochromatic images and material density maps, and images can be post-processed. It is also indicated for lung cancer screening for patients who meet the inclusion criteria of published governmental or professional medical society programs or protocols.
- Compared with (the predicate)
- K213715, Revolution Apex
- How it differs
- The summary says the Photonova Spectra uses a new silicon photon counting detector in place of the predicate's energy integrating detector with gemstone scintillator. It also adds an extra small focal spot, a new deep learning reconstruction called TrueFidelity DL for PCCT, a 1024 x 1024 image matrix option, and a 1.0 to 2.0 second rotation option.
- Testing the summary lists
- Electrical safety (AAMI/ANSI ES 60601-1, IEC 60601-1 Ed. 3.2, and collateral and particular standards)
- Radiation and X-ray system standards (21 CFR Subchapter J, NEMA XR 25 and XR 28)
- Software verification and validation, risk management, and bench testing
- Image quality and dose performance testing, including low contrast detectability studies and IEC 61223-3-5 performance elements
- Simulated use testing (validation)
- Clinical reader study with US board-certified radiologists reviewing images across Neuro, Body, and Cardiac/Chest scenarios
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K253520 took 128 calendar days from receipt to decision. Of the latest 24 other decisions in product code JAK, 15 took longer, 9 were faster; their median was 168 days.
Latest decisions in JAK
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260391 | AVIEW Aorta MetricsCoreline Soft Co., Ltd. · Traditional | Oct 16, 2026252 days |
| K261225 | CAS-One IRCascination AG · Traditional | Sep 29, 2026167 days |
| K261125 | BodyTom 64 (NL4100); OmniTom Elite (NL5000)Samsung HME America, Inc., dba NeuroLogica Corp. · Traditional | Sep 22, 2026169 days |
| K261229 | SCENARIA View Phase 6.0Fujifilm Corporation · Traditional | Sep 18, 2026156 days |
| K253836 | FCT iStream Phase 2Fujifilm Corporation · Traditional | Sep 4, 2026277 days |
| K253450 | BodyTom Elite (NL4000)NeuroLogica Corp., a Samsung Company · Traditional | Jul 28, 2026299 days |
| K253770 | CardIQ SuiteGE Medical Systems SCS · Traditional | Jul 21, 2026237 days |
| K253560 | Enhanced Boundary for PCCTGe Medical Systems, LLC · Traditional | Jun 24, 2026219 days |
| K253026 | Respiratory Gating for Scanners (2.1)Varian Medical Systems, Inc. · Traditional | Jun 4, 2026258 days |
| K253596 | Aquilion ONE (TSX-308A/TSX-306A) V2.0Canon Medical Systems Corporation · Traditional | May 20, 2026183 days |
Other 510(k)s from Ge Medical Systems, LLC
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253560 | Enhanced Boundary for PCCTGe Medical Systems, LLC · Traditional | Jun 24, 2026219 days |
| K253502 | Critical Care Suite with Enteric Tube Positioning AI AlgorithmGe Medical Systems, LLC · Traditional | Apr 14, 2026166 days |
| K253780 | SIGNA™ BoltGe Medical Systems, LLC · Traditional | Feb 6, 202672 days |
| K253779 | SIGNA™ Sprint SelectGe Medical Systems, LLC · Traditional | Feb 5, 202671 days |
| K252379 | AIR Recon DLGe Medical Systems, LLC · Traditional | Dec 23, 2025146 days |
| K251399 | SIGNA™ SprintGe Medical Systems, LLC · Traditional | Sep 11, 2025128 days |
| K250941 | Revolution VibeGe Medical Systems, LLC · Traditional | Aug 1, 2025126 days |
| K243667 | Sonic DLGe Medical Systems, LLC · Traditional | Jun 5, 2025190 days |
| K250379 | SIGNA Prime EliteGe Medical Systems, LLC · Traditional | Mar 17, 202534 days |
| K241242 | SIGNA MAGNUSGe Medical Systems, LLC · Traditional | Oct 17, 2024167 days |
| K233750 | ECG-less CardiacGe Medical Systems, LLC · Traditional | Apr 2, 2024132 days |
| K223523 | Sonic DLGe Medical Systems,Llc (Ge Healthcare) · Traditional | May 30, 2023188 days |
Watch JAK in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get JAK every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in JAK. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K253520Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:JAKRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:JAKRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:JAK AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Ge Medical Systems, LLC"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil