FDA device news · Product code JAK

FDA 510(k) clearance

Photonova Spectra, Photonova Spectra Select (K253520): Ge Medical Systems, LLC 510(k) clearance

Waukesha, WI

Follow JAK: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Mar 20, 2026
Received
Nov 12, 2025
510(k) type
Traditional
Product code
JAKSystem, X-Ray, Tomography, Computed
Review panel
Radiology
Device class
Class II
Regulation
21 CFR 892.1750
Third-party review
No
Summary or statement
Summary
Review time
128 daysreceipt to decision, calendar days
Median in JAK
168 daysthe latest 24 other decisions
JAK decisions
21in the last 12 months
Applicant’s 510(k)s
113in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
Photonova Spectra and Photonova Spectra Select are X-ray computed tomography (CT) scanners that create cross-sectional images of the body by computer reconstruction of X-ray transmission data. They use a new silicon-based photon counting detector that directly converts X-ray photons into electrical signals and sorts them into 8 energy bins, so each scan captures multi-energy data.
What it’s cleared for
The system is cleared to produce cross-sectional body images for head, whole body, cardiac, and vascular CT applications, for patients of all ages. It natively generates high resolution monochromatic images and material density maps, and images can be post-processed. It is also indicated for lung cancer screening for patients who meet the inclusion criteria of published governmental or professional medical society programs or protocols.
Compared with (the predicate)
How it differs
The summary says the Photonova Spectra uses a new silicon photon counting detector in place of the predicate's energy integrating detector with gemstone scintillator. It also adds an extra small focal spot, a new deep learning reconstruction called TrueFidelity DL for PCCT, a 1024 x 1024 image matrix option, and a 1.0 to 2.0 second rotation option.
Testing the summary lists
  • Electrical safety (AAMI/ANSI ES 60601-1, IEC 60601-1 Ed. 3.2, and collateral and particular standards)
  • Radiation and X-ray system standards (21 CFR Subchapter J, NEMA XR 25 and XR 28)
  • Software verification and validation, risk management, and bench testing
  • Image quality and dose performance testing, including low contrast detectability studies and IEC 61223-3-5 performance elements
  • Simulated use testing (validation)
  • Clinical reader study with US board-certified radiologists reviewing images across Neuro, Body, and Cardiac/Chest scenarios

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K253520 took 128 calendar days from receipt to decision. Of the latest 24 other decisions in product code JAK, 15 took longer, 9 were faster; their median was 168 days.

Latest decisions in JAK

NumberDevice and applicantDecision and review time
K260391 AVIEW Aorta MetricsCoreline Soft Co., Ltd. · Traditional Oct 16, 2026252 days
K261225 CAS-One IRCascination AG · Traditional Sep 29, 2026167 days
K261125 BodyTom 64 (NL4100); OmniTom Elite (NL5000)Samsung HME America, Inc., dba NeuroLogica Corp. · Traditional Sep 22, 2026169 days
K261229 SCENARIA View Phase 6.0Fujifilm Corporation · Traditional Sep 18, 2026156 days
K253836 FCT iStream Phase 2Fujifilm Corporation · Traditional Sep 4, 2026277 days
K253450 BodyTom Elite (NL4000)NeuroLogica Corp., a Samsung Company · Traditional Jul 28, 2026299 days
K253770 CardIQ SuiteGE Medical Systems SCS · Traditional Jul 21, 2026237 days
K253560 Enhanced Boundary for PCCTGe Medical Systems, LLC · Traditional Jun 24, 2026219 days
K253026 Respiratory Gating for Scanners (2.1)Varian Medical Systems, Inc. · Traditional Jun 4, 2026258 days
K253596 Aquilion ONE (TSX-308A/TSX-306A) V2.0Canon Medical Systems Corporation · Traditional May 20, 2026183 days

Other 510(k)s from Ge Medical Systems, LLC

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K253560 Enhanced Boundary for PCCTGe Medical Systems, LLC · Traditional Jun 24, 2026219 days
K253502 Critical Care Suite with Enteric Tube Positioning AI AlgorithmGe Medical Systems, LLC · Traditional Apr 14, 2026166 days
K253780 SIGNA™ BoltGe Medical Systems, LLC · Traditional Feb 6, 202672 days
K253779 SIGNA™ Sprint SelectGe Medical Systems, LLC · Traditional Feb 5, 202671 days
K252379 AIR Recon DLGe Medical Systems, LLC · Traditional Dec 23, 2025146 days
K251399 SIGNA™ SprintGe Medical Systems, LLC · Traditional Sep 11, 2025128 days
K250941 Revolution VibeGe Medical Systems, LLC · Traditional Aug 1, 2025126 days
K243667 Sonic DLGe Medical Systems, LLC · Traditional Jun 5, 2025190 days
K250379 SIGNA Prime EliteGe Medical Systems, LLC · Traditional Mar 17, 202534 days
K241242 SIGNA MAGNUSGe Medical Systems, LLC · Traditional Oct 17, 2024167 days
K233750 ECG-less CardiacGe Medical Systems, LLC · Traditional Apr 2, 2024132 days
K223523 Sonic DLGe Medical Systems,Llc (Ge Healthcare) · Traditional May 30, 2023188 days

Watch JAK in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get JAK every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in JAK. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.