FDA device news · Product code CEM

FDA 510(k) clearance

ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas pro integrated solutions (K253491): Roche Diagnostics 510(k) clearance

Indianapolis, IN

Follow CEM: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Feb 12, 2026
Received
Oct 24, 2025
510(k) type
Traditional
Product code
CEMElectrode, Ion Specific, Potassium
Review panel
Clinical Chemistry
Device class
Class II
Regulation
21 CFR 862.1600
Third-party review
No
Summary or statement
Summary
Review time
111 daysreceipt to decision, calendar days
Median in CEM
135 daysthe latest 24 other decisions
CEM decisions
1in the last 12 months
Applicant’s 510(k)s
488in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The cobas pro integrated solutions is an automated lab analyzer. Its ISE (ion-selective electrode) unit measures sodium, potassium and chloride in serum, plasma or urine. An ion-selective electrode uses a special membrane that produces a small electrical voltage only from the ion being measured, and that voltage is used to calculate the ion's concentration.
What it’s cleared for
The ISE unit is cleared for the quantitative measurement of sodium, potassium and chloride in serum, plasma or urine. The summary says these measurements are used in diagnosing and treating conditions involving electrolyte imbalance, such as aldosteronism, diabetes insipidus, dehydration, potassium-related disorders, cystic fibrosis and diabetic acidosis.
Compared with (the predicate)
  • K060373, cobas c 501
  • K191899, ISE indirect Na for Gen.2 on cobas pro ISE
How it differs
The summary says the cobas ISE neo runs faster, at 900 tests per hour compared with 600, and uses a larger default sample volume of 15 µL compared with 9.7 µL. It is also self-contained, with its own temperature control, vacuum and water supply rather than sharing these with an adjacent analyzer module.

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K253491 took 111 calendar days from receipt to decision. Of the latest 24 other decisions in product code CEM, 14 took longer, 10 were faster; their median was 135 days.

Latest decisions in CEM

NumberDevice and applicantDecision and review time
K253491 ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas pro integrated solutionsRoche Diagnostics · Traditional Feb 12, 2026111 days
K241037 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · Traditional Jan 14, 2025273 days
K220396 EasyStat 300Medica Corporation · Traditional Sep 26, 2023592 days
K230890 ISE ElectrodesRandox Laboratories, Ltd. · Special Sep 8, 2023161 days
K200544 SmartLyte® Plus Electrolyte Analyzer Na+/K+/Cl-/Ca++/Li+Diamond Diagnostics, Inc. · Traditional Oct 1, 2020212 days
K121012 FLEXLAB 3.6, ACCELERATOR A3600Inpeco S.P.A. · Traditional Aug 31, 2012150 days
K121040 SMARTLYTE ELECTROLYTE ANALYZERDiamond Diagnostics, Inc. · Traditional Aug 30, 2012147 days
K112995 GEM PREMIER 4000Instrumentation Laboratory CO · Traditional May 3, 2012209 days
K110520 BIOLIS 12ITokyo Boeki Medisys, Inc. · Traditional Mar 23, 2012394 days
K112412 AU5800(R) CHEMISTRY ANALYZERBeckman Coulter, Inc. · Traditional Dec 23, 2011123 days

Other 510(k)s from Roche Diagnostics

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
DEN260027 Elecsys HBsAg II quant II RxDxRoche Diagnostics · Direct Oct 20, 2026140 days
K260221 Elecsys EBV EBNA IgGRoche Diagnostics · Traditional Oct 2, 2026249 days
K260232 Elecsys EBV VCA IgGRoche Diagnostics · Traditional Oct 2, 2026249 days
K260233 Elecsys EBV IgMRoche Diagnostics · Traditional Oct 2, 2026249 days
K261686 Elecsys Phospho-Tau (217P) PlasmaRoche Diagnostics · Traditional Aug 19, 202690 days
K252323 cobas pulse blood glucose monitoring systemRoche Diagnostics · Dual Track Jun 12, 2026322 days
K252431 Elecsys CalcitoninRoche Diagnostics GmbH · Traditional Apr 24, 2026266 days
K252280 Elecsys Anti-SARS-CoV-2 SRoche Diagnostics · Traditional Apr 13, 2026265 days
K260046 Elecsys Anti-HBc IIRoche Diagnostics · Traditional Apr 7, 202690 days
K260048 Elecsys Anti-HAV IgMRoche Diagnostics · Traditional Apr 7, 202690 days
K260049 Elecsys Anti-HBc IgMRoche Diagnostics · Traditional Apr 7, 202690 days
K260026 Tina-quant Cardiac high sensitivity CRP IIIRoche Diagnostics · Traditional Apr 2, 202687 days

Watch CEM in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get CEM every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in CEM. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.