FDA device news · Product code QDG

FDA 510(k) clearance

Dendrite Imaging System (K253303): Dendrite Imaging, Inc. 510(k) clearance

Berkeley, CA

Follow QDG: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Nov 28, 2025
Received
Sep 29, 2025
510(k) type
Traditional
Product code
QDGParathyroid Autofluorescence Imaging Device
Review panel
General, Plastic Surgery
Device class
Class II
Regulation
21 CFR 878.4550
Third-party review
No
Summary or statement
Summary
Review time
60 daysreceipt to decision, calendar days
Median in QDG
117 daysthe latest 5 other decisions
QDG decisions
0in the last 12 months
Applicant’s 510(k)s
1in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Dendrite Imaging System is a handheld camera head with a near ultraviolet (NUV) LED light source and a touchscreen display. Nerve tissue naturally gives off a faint glow, called autofluorescence, when NUV light hits it. The camera captures this glow and shows it on the screen in real time, and the surgeon can save images or video to a USB drive.
What it’s cleared for
The device is cleared for real-time capturing and viewing of autofluorescent images of nerves during surgery, as an adjunct to help locate nerves. It is meant to assist, not replace, the surgeon's visual assessment. It must not be used to confirm that nerve tissue is absent, and it should only be used to help find nerves that have already been seen visually.
Compared with (the predicate)
How it differs
The Dendrite Imaging System uses a 385 nm LED light to excite tissue, while the Fluobeam LX uses a 750 nm laser. It images nerve tissue, while the predicate images the parathyroid gland.
Testing the summary lists
  • Electrical safety and electromagnetic compatibility (IEC 60601-1, IEC 60601-1-6, IEC 60601-1-2, IEC TS 60601-4-2)
  • Light safety (IEC 62471)
  • Packaging validation after conditioning and shipping simulation (ISTA-3A)
  • Software verification and lifecycle process (IEC 62304 principles)
  • Cybersecurity vulnerability and penetration testing
  • Animal testing in guinea pigs under GLP rules (21 CFR Part 58) and an observational clinical study of 23 subjects in open surgery

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K253303 took 60 calendar days from receipt to decision. Of the latest 5 other decisions in product code QDG, 3 took longer, 2 were faster; their median was 117 days.

Latest decisions in QDG

NumberDevice and applicantDecision and review time
K253303 Dendrite Imaging SystemDendrite Imaging, Inc. · Traditional Nov 28, 202560 days
K250455 FLUOBEAM LX Imaging System (FBLX); FLUOBEAM LM Imaging System (FBLM)Fluoptics Sas (A Getinge Group Company} · Traditional Apr 17, 202558 days
K233564 FLUOBEAM® LX Imaging System (FB-LX); FLUOBEAM® LX Red Imaging System (FB-LXR)Fluoptics Sas (A Getinge Group Company) · Special Dec 15, 202339 days
K230898 FLUOBEAM® LX RedFluoptics Sas · Special Jul 28, 2023119 days
K190891 Fluobeam LXFluoptics · Traditional Jul 31, 2019117 days
DEN170092 Fluobeam 800 Clinic Imaging Device used with Fluocase 800 Control SystemFluoptics · Direct Nov 2, 2018315 days

Other 510(k)s from Dendrite Imaging, Inc.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

This is the only 510(k) in openFDA under this applicant name.

Watch QDG in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get QDG every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in QDG. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.