FDA device news · Product code MMK

FDA 510(k) clearance

Gongdong Sharps Container (SCR-01Q, SCR-01G, SCR-03G) (K253222): Zhejiang Gongdong Medical Technology Co., Ltd. 510(k) clearance

Taizhou, CN

Follow MMK: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Dec 22, 2025
Received
Sep 29, 2025
510(k) type
Traditional
Product code
MMKContainer, Sharps
Review panel
General Hospital
Device class
Class II
Regulation
21 CFR 880.5570
Third-party review
No
Summary or statement
Summary
Review time
84 daysreceipt to decision, calendar days
Median in MMK
247 daysthe latest 24 other decisions
MMK decisions
3in the last 12 months
Applicant’s 510(k)s
3in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Gongdong Sharps Container (models SCR-01Q, SCR-01G and SCR-03G) is a disposable plastic box used to collect used needles, syringes, lancets and similar sharp items. It is made of polypropylene and is designed to resist punctures, leaks, impact and tipping. It is non-sterile and is not meant to touch the patient.
What it’s cleared for
The containers are cleared to hold used, contaminated medical sharps and stay closed during transport to final disposal. They are intended for qualified personnel in health care facilities and other facilities where medical sharps may be used, and not for areas with unsupervised patient access.
Compared with (the predicate)
  • K222905, Maxcon Sharps Container (MA1112)
  • K180984, Maxcon Sharps Container (MA1221)
How it differs
The summary says the dimensions and weights differ from the predicate devices. It also says the containers are similar in size and accessories, and that the differences do not impair their intended function of disposing of and storing sharps waste.
Testing the summary lists
  • Container stability, handle strength, penetration resistance, and damage and leakage after dropping and toppling (ISO 23907-1:2019)
  • Stacking test (49 CFR 178.606)
  • Vibration test (49 CFR 178.608)
  • Sharps access and closure for repeated openings and closings (manufacturer internal SOP)
  • Usable capacity test (manufacturer internal SOP)
  • Leak proof on sides and bottom (29 CFR 1910.1030)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K253222 took 84 calendar days from receipt to decision. Of the latest 24 other decisions in product code MMK, 22 took longer, 2 were faster; their median was 247 days.

Latest decisions in MMK

NumberDevice and applicantDecision and review time
K260845 Huahong Sharps Container (HHR-I, HHR-II)Tianjin Huahong Technology Co., Ltd. · Traditional Aug 5, 2026142 days
K253773 Guanhong Sharps Container (GHW-1F1, GHW-1F4, GH-6FA, GHW-2FR, GHW-8FR, GH-3FS, GHW-15FS, GHW-27FS, GH-5FSW, GHW-3FT, GHW-4FT, GHW-5FT, GHW-8FT); GuanhongTaizhou Huangyan Guanhong Plastic Steel Products Factory · Traditional May 22, 2026177 days
K253222 Gongdong Sharps Container (SCR-01Q, SCR-01G, SCR-03G)Zhejiang Gongdong Medical Technology Co., Ltd. · Traditional Dec 22, 202584 days
K251874 PureWay Sharps Container 1-Gal, 2-Gal, 3-Gal (800011 800012, 800013)Pureway Compliance, Inc. · Traditional Oct 14, 2025118 days
K252637 Community Containers (Flap and Daisy)Keter Canada, Inc. · Traditional Oct 14, 202555 days
K231484 PureWay 1.4 Quart Sharps CollectorPureway Compliance, Inc. · Traditional Sep 14, 2023114 days
K222906 Maxcon Reusable Sharps Container, 17 Gallon Sharps Container MA1421Ningbo Maxcon Medical Technology Co., Ltd. · Traditional Jun 16, 2023266 days
K222905 Maxcon Sharps Container (1 QT Sharps Container, MA1112), Maxcon Sharps Container (5.4 QT Sharps Container, MA1212), Maxcon Sharps Container (5.4 QT Sharps Container, MA1213), Maxcon Sharps Container (7 Liter Sharps Container, MA1324)Ningbo Maxcon Medical Technology Co., Ltd. · Traditional Jun 1, 2023251 days
K213274 Qlicksmart BladeFlask UNIVERSALQlicksmart Pty , Ltd. · Abbreviated Jun 22, 2022265 days
K211890 Promisemed Sharps containerPromisemed Hangzhou Meditech Co., Ltd. · Traditional Nov 17, 2021149 days

Other 510(k)s from Zhejiang Gongdong Medical Technology Co., Ltd.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K220739 Disposable Urine Collection TubeZhejiang Gongdong Medical Technology Co., Ltd. · Traditional Nov 17, 2022248 days
K201523 Sharps containerZhejiang Gongdong Medical Technology Co., Ltd. · Traditional Nov 12, 2020157 days

Watch MMK in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get MMK every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in MMK. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.