FDA device news · Product code SAH
FDA 510(k) clearance
GT Metabolic MagDI System (MAG-01, MAG-02, DS-01) (K253102): Gt Metabolic Solutions, Inc. 510(k) clearance
St. Michael, MN
- Decision
- Substantially Equivalent
- Decision date
- Jun 4, 2026
- Received
- Sep 24, 2025
- 510(k) type
- Traditional
- Product code
- SAHMagnetic Compression Anastomosis System
- Review panel
- General, Plastic Surgery
- Device class
- Class II
- Regulation
- 21 CFR 878.4816
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 253 daysreceipt to decision, calendar days
- Median in SAH
- 122 daysthe latest 9 other decisions
- SAH decisions
- 4in the last 12 months
- Applicant’s 510(k)s
- 5in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The GT Metabolic MagDI System is a magnetic compression anastomosis system. It uses a set of two sterile, single-use magnets (40mm for MAG-01 or 50mm for MAG-02) to join two sections of bowel. According to the summary, the magnets compress the tissue between them, causing necrosis (tissue death) over about 7 to 21 days. The device and the compressed tissue then detach and are passed with the stool.
- What it’s cleared for
- The device is cleared to create side-to-side anastomoses (surgical connections between sections of bowel) throughout the small bowel during minimally invasive and laparoscopic surgery, for adult patients over 21. Once wound strength is sufficient to hold the connection, the device is passed from the body. The summary states that the effects of the device on weight loss were not studied.
- How it differs
- The summary says the device is the same as the predicate in all technological characteristics and has the same intended use. The one change is an expanded indication, from duodeno-ileal anastomosis to anastomosis throughout the small bowel, based on clinical testing.
- Testing the summary lists
- Preclinical testing: not required, based on risk analysis
- Animal testing: not required, based on risk analysis
- Real-world clinical use review (26 cases using 40mm and 50mm magnets)
- Ongoing post-market surveillance study on internal hernia and bowel obstruction
- Labeling conformance review (21 CFR 801 and ISO 15223)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K253102 took 253 calendar days from receipt to decision. Of the latest 9 other decisions in product code SAH, 9 were faster; their median was 122 days.
Latest decisions in SAH
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261630 | GI Windows Flexagon Plus OTOLoc SystemGI Windows Surgical · Traditional | Oct 28, 2026163 days |
| K261467 | GT Metabolic Magnet System (MAG-04, DS-01)GT Metabolic Magnet System (MAG-05, DS-01)Gt Metabolic Solutions, Inc. · Traditional | Sep 16, 2026135 days |
| K253102 | GT Metabolic MagDI System (MAG-01, MAG-02, DS-01)Gt Metabolic Solutions, Inc. · Traditional | Jun 4, 2026253 days |
| K253550 | Self-Forming Magnet (Flexagon Plus OTOLoc)GI Windows, Inc. · Traditional | Jan 9, 202656 days |
| K250541 | Self-Forming Magnet (Flexagon Plus OTOLoc)GI Windows, Inc. · Traditional | Aug 1, 2025158 days |
| K243482 | Self-Forming Magnet (Flexagon)GI Windows, Inc. · Traditional | Jun 11, 2025215 days |
| K243359 | MagDI System (MAG-02, DS-01)Gt Metabolic Solutions, Inc. · Traditional | Feb 28, 2025122 days |
| K243213 | Self-Forming Magnet (FLEX SFM)GI Windows, Inc. · Traditional | Jan 29, 2025118 days |
| K242086 | MagDI System (MAG-01, DS-01)Gt Metabolic Solutions, Inc. · Traditional | Oct 24, 202499 days |
| DEN240013 | MagDI SystemGt Metabolic Solutions, Inc. · Post-NSE | Jul 2, 202498 days |
Other 510(k)s from Gt Metabolic Solutions, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261467 | GT Metabolic Magnet System (MAG-04, DS-01)GT Metabolic Magnet System (MAG-05, DS-01)Gt Metabolic Solutions, Inc. · Traditional | Sep 16, 2026135 days |
| K243359 | MagDI System (MAG-02, DS-01)Gt Metabolic Solutions, Inc. · Traditional | Feb 28, 2025122 days |
| K242086 | MagDI System (MAG-01, DS-01)Gt Metabolic Solutions, Inc. · Traditional | Oct 24, 202499 days |
| DEN240013 | MagDI SystemGt Metabolic Solutions, Inc. · Post-NSE | Jul 2, 202498 days |
Watch SAH in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get SAH every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in SAH. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K253102Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:SAHRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:SAHRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:SAH AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Gt Metabolic Solutions, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.