FDA device news · Product code QCF

FDA 510(k) clearance

Portable Neuromodulation Stimulator (PoNS) (K253061): Helius Medical, Inc. 510(k) clearance

Newton, PA

Follow QCF: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
May 13, 2026
Received
Sep 22, 2025
510(k) type
Traditional
Product code
QCFElectrical Tongue Stimulator To Treat Motor Deficits
Review panel
Neurology
Device class
Class II
Regulation
21 CFR 882.5889
Third-party review
No
Summary or statement
Summary
Review time
233 daysreceipt to decision, calendar days
Median in QCF
234 daysthe latest 1 other decisions
QCF decisions
1in the last 12 months
Applicant’s 510(k)s
2in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Portable Neuromodulation Stimulator (PoNS) is a tongue stimulator. It has a Controller worn around the neck and a Mouthpiece placed in the mouth with 143 gold-plated electrodes that deliver mild electrotactile stimulation (a small electrical signal felt as vibration, tingle, or pressure) to the top surface of the tongue. The system also includes a charger, software, and carry and retainer cases.
What it’s cleared for
The PoNS is cleared as a treatment for dynamic gait deficit (trouble walking that involves changing movement or balance) caused by chronic stroke symptoms. It is for patients 22 years of age and older, used with a supervised therapeutic exercise program, and is available by prescription only.
Compared with (the predicate)
  • DEN200050, Portable Neuromodulation Stimulator (PoNS TM)
How it differs
The summary says the device is identical to the predicate in design, materials, energy source, and technology. The main difference is the indication: the predicate was indicated as a short term treatment of gait deficit from mild to moderate symptoms of multiple sclerosis, while this device is indicated for dynamic gait deficit from chronic stroke symptoms.
Testing the summary lists
  • Shelf life (4-year Mouthpiece shelf life; 3-year Controller use life)
  • Electrical safety and electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10)
  • Software documentation (Basic Documentation level) and cybersecurity documentation with cyber threat analysis
  • Human factors/usability (IEC 60601-1-6, BS EN 62366-1, ISO 14971)
  • Biocompatibility (leveraged from predicate DEN200050, not separately tested)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K253061 took 233 calendar days from receipt to decision. Of the latest 1 other decision in product code QCF, 1 took longer; their median was 234 days.

Latest decisions in QCF

NumberDevice and applicantDecision and review time
K253061 Portable Neuromodulation Stimulator (PoNS)Helius Medical, Inc. · Traditional May 13, 2026233 days
DEN200050 Portable Neuromodulation Stimulator (PoNS)Helius Medical, Inc. · Direct Mar 26, 2021234 days

Other 510(k)s from Helius Medical, Inc.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
DEN200050 Portable Neuromodulation Stimulator (PoNS)Helius Medical, Inc. · Direct Mar 26, 2021234 days

Watch QCF in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get QCF every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in QCF. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.