FDA device news · Product code EZD
FDA 510(k) clearance
GentleCath Air for Men; GentleCath Glide Intermittent Catheter; GentleCath Air for Women (K252943): Convatec Limited 510(k) clearance
Deeside, Flintshire, GB
- Decision
- Substantially Equivalent
- Decision date
- Apr 1, 2026
- Received
- Sep 15, 2025
- 510(k) type
- Traditional
- Product code
- EZDCatheter, Straight
- Review panel
- Gastroenterology, Urology
- Device class
- Class II
- Regulation
- 21 CFR 876.5130
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 198 daysreceipt to decision, calendar days
- Median in EZD
- 141 daysthe latest 24 other decisions
- EZD decisions
- 12in the last 12 months
- Applicant’s 510(k)s
- 11in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- GentleCath Air for Men, GentleCath Glide Intermittent Catheter, and GentleCath Air for Women are sterile, single-use, hydrophilic intermittent urinary catheters. A catheter is a thin hollow tube inserted through the urethra to drain urine from the bladder. Hydrophilic means the surface becomes slippery when wetted, and the summary describes a surface technology called FeelClean Technology used on these catheters.
- What it’s cleared for
- The devices are cleared for intermittent drainage of the urinary bladder in people who need help draining their bladder because of urinary retention or urinary system dysfunction. The GentleCath Air for Men and GentleCath Air for Women are for adults and females, respectively, and the GentleCath Glide is for adults and pediatric patients.
- Compared with (the predicate)
- How it differs
- The summary lists several differences from the predicates. The Glide's sterilization changed from ethylene oxide to X-ray, and the Air for Women funnel's HDPE raw material lacks a trace additive found in its predicate. The eyelet tolerance for some sizes changed from plus or minus 0.2 mm to plus or minus 0.3 mm.
- Testing the summary lists
- In vitro and ex vivo coefficient of friction testing and insertion/withdrawal force testing (friction force and lubricity claims)
- Post-market study with patient comfort questionnaires (comfort claim)
- Sterilization validation for the proposed X-ray sterilization method change
- Prolonged stability testing to support the proposed shelf life extension
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K252943 took 198 calendar days from receipt to decision. Of the latest 24 other decisions in product code EZD, 7 took longer, 17 were faster; their median was 141 days.
Latest decisions in EZD
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K263512 | Luja femaleColoplast Corp. · Special | Oct 28, 202629 days |
| K261418 | Liquick X-treme, Liquick X-treme PlusTeleflex Medical · Traditional | Sep 23, 2026146 days |
| K260737 | Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Female (CH06 - CH16); Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Straight (CH06 - CH18); Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Coude (CH10 - CH18).Convatec · Traditional | Aug 5, 2026152 days |
| K261658 | LoFric Origo Pro CoudéWellspect AB · Special | Jul 17, 202659 days |
| K262045 | Luja SoftColoplast A.S. · Special | Jul 7, 202620 days |
| K260394 | Hydrophilic coated intermittent nalaton catheter for single use-ready to use (NCR)Hangzhou Jimushi Meditech Co., Ltd. · Traditional | Jun 24, 2026138 days |
| K253295 | Coude Sleeved IC; Coude Plus Sleeved ICHollister Incorporated · Traditional | May 18, 2026231 days |
| K260999 | LoFric Elle ProWellspect AB · Special | Apr 24, 202629 days |
| K252943 | GentleCath Air for Men; GentleCath Glide Intermittent Catheter; GentleCath Air for WomenConvatec Limited · Traditional | Apr 1, 2026198 days |
| K260502 | Liv Pre-lubricated Intermittent CatheterHR Healthcare · Traditional | Mar 12, 202627 days |
Other 510(k)s from Convatec Limited
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253263 | ConvaVAC™ Negative Pressure Wound Therapy SystemConvatec Limited, Gdc Building · Traditional | Jun 24, 2026268 days |
| K250891 | Cure Catheter Insertion Kit (K1, K2, K2-90, K3); Cure Hydrophilic Catheter Kit (HM12UK, HM14UK, HM16UK); Cure Pocket Catheter Kit (M14UK); Cure Catheter Closed System Kit (CS8, CS10, CS12, CS14); Cure Catheter Closed System Kit (CS14CConvatec Limited · Special | Jun 6, 202573 days |
| K213283 | GentleCath Air for Men Hydrophilic Intermittent Urinary CatheterConvatec Limited · Traditional | Jun 22, 2022264 days |
| K190114 | Flexi-Seal PROTECT PLUS Fecal Management System (with ENFit Connector)Convatec Limited · Traditional | Apr 30, 201996 days |
| K180205 | AVELLE Negative Pressure Wound Therapy SystemConvatec Limited · Traditional | Oct 19, 2018268 days |
| K181206 | GentleCath Glide Intermittent Urinary CatheterConvatec Limited · Traditional | Aug 16, 2018101 days |
| K173675 | AQUACEL Ag+ EXTRA Enhanced Hydrofiber Dressing with Silver and Strengthening FiberConvatec Limited · Traditional | Jul 20, 2018232 days |
| K172924 | GentleCath Intermittent Urinary CatheterConvatec Limited · Traditional | Nov 21, 201757 days |
| K161344 | GentleCath Glide Intermittent CatheterConvatec Limited · Traditional | Sep 20, 2016130 days |
| K152926 | AQUACEL Ag Surgical SP DressingConvatec Limited · Special | Jan 21, 2016108 days |
Watch EZD in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get EZD every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in EZD. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K252943Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:EZDRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:EZDRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:EZD AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Convatec Limited"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil