FDA device news · Product code FMI
FDA 510(k) clearance
Pen Needle (K252886): Hh Global Technology, Inc. 510(k) clearance
Buford, GA
- Decision
- Substantially Equivalent
- Decision date
- Jan 6, 2026
- Received
- Sep 10, 2025
- 510(k) type
- Traditional
- Product code
- FMINeedle, Hypodermic, Single Lumen
- Review panel
- General Hospital
- Device class
- Class II
- Regulation
- 21 CFR 880.5570
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 118 daysreceipt to decision, calendar days
- Median in FMI
- 120 daysthe latest 24 other decisions
- FMI decisions
- 12in the last 12 months
- Applicant’s 510(k)s
- 2in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- A sterile, single-use pen needle that screws onto a pen injector to inject drug under the skin (subcutaneous injection). It is offered in ordinary and safety versions, with a needle shield and, for safety types, a needle cover and spring that help protect against accidental needle sticks after use. Each needle comes individually packaged and sterilized by irradiation.
- What it’s cleared for
- The Indications for Use state that the Pen Needles are sterile, single-use needles intended for use with pen injector devices to inject drug. The summary lists both prescription and over-the-counter use, and says the device is intended for healthcare personnel, adult patients, and adult lay persons.
- How it differs
- The summary says the subject device differs from predicate K220129 in trade name, over-the-counter versus prescription use labeling, sterilization method (irradiation instead of ethylene oxide), needle gauges (33G and 34G added), needle lengths (3.5mm, 10mm and 12mm added), and materials.
- Testing the summary lists
- Biocompatibility (ISO 10993-1, ISO 10993 series: cytotoxicity, skin irritation, sensitization, systemic toxicity, hemolysis, pyrogenicity)
- Performance (ISO 9626, ISO 7864, ISO 11608-2)
- Sterility and packaging (ISO 11137-2 radiation dose validation, USP <85> endotoxin, ASTM F88, F1929, F1886)
- Shelf life (ASTM F1980-21 accelerated aging, 5 years)
- Sharps injury protection (simulated clinical use study on 600 safety-configuration samples per ISO 23908:2011 and FDA sharps guidance)
- Compatibility with pen injectors (ISO 11608-1 compliance claimed)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K252886 took 118 calendar days from receipt to decision. Of the latest 24 other decisions in product code FMI, 13 took longer, 11 were faster; their median was 120 days.
Latest decisions in FMI
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261266 | MOWANA™ Auto DuoPENTAMASTER MEDIQ SDN BHD · Traditional | Sep 17, 2026154 days |
| K261719 | Promisemed X-Safety Pen NeedlePromisemed Hangzhou Meditech Co., Ltd. · Special | Aug 25, 202691 days |
| K262427 | Pen Needle (Ordinary Type I, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Safety Type VII, Ordinary Type IB, Safety type VIII, Safety Type IX, Safety Type X, Ordinary Type XI, Safety Type XII)Hh Global Technology, Inc. · Special | Aug 10, 202625 days |
| K260933 | Safety Sterile Hypodermic Needles (NR-S1816R, NR-S1916R, NR-S1925R, NR-S1938R, NR-S2016R, NR-S2025R, NR-S2038R, NR-S2051R, NR-S2116R, NR-S2125R, NR-S2138R, NR-S2151R, NR-S2216R, NR-S2225R, NR-S2238R, NR-S2251R, NR-S2316R, NR-S2325R, NR-S2338R, NR-S2351R, NR-S2516R, NR-S2525R, NR-S2538R, NR-S2551R, NR-S2516T, NR-S2525T, NR-S2538T, NR-S2551T)Hangzhou Qiantang Longyue Biotechnology Co., Ltd. · Traditional | Aug 7, 2026140 days |
| K261829 | Profoject™ Dual-Safety Pen NeedlesCMT Health PTE., Ltd. · Traditional | Jul 29, 202657 days |
| K260739 | Profoject™ Pen Needles (Model A; Model B)CMT Health PTE., Ltd. · Traditional | Jul 15, 2026131 days |
| K260855 | Profoject™ Safety Pen NeedlesCMT Health PTE., Ltd. · Traditional | Jun 10, 202686 days |
| K254273 | Verisafe Safety Sterile Needles (SSN)Promisemed Hangzhou Meditech Co., Ltd. · Special | Apr 16, 2026107 days |
| K252908 | Easydrip Classic Pen Needle; Easydrip Plus Pen Needle; Easydrip Classic Pro Pen Needle; Easydrip Plus Pro Pen NeedleSandstone Medical (Suzhou), Inc. · Special | Mar 25, 2026194 days |
| K253047 | PRESSONE™Nipro Medical Corporation · Traditional | Mar 24, 2026183 days |
Other 510(k)s from Hh Global Technology, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262427 | Pen Needle (Ordinary Type I, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Safety Type VII, Ordinary Type IB, Safety type VIII, Safety Type IX, Safety Type X, Ordinary Type XI, Safety Type XII)Hh Global Technology, Inc. · Special | Aug 10, 202625 days |
Watch FMI in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get FMI every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in FMI. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K252886Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:FMIRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:FMIRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:FMI AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Hh Global Technology, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil