FDA device news · Product code QRC

FDA 510(k) clearance

in2 Smart (in2S) (K252809): Virility Medical , Ltd. 510(k) clearance

Hod Hasharon, IL

Follow QRC: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Apr 6, 2026
Received
Sep 3, 2025
510(k) type
Traditional
Product code
QRCNon-Implanted Electrical Stimulation Device For Management Of Premature Ejaculation
Review panel
Gastroenterology, Urology
Device class
Class II
Regulation
21 CFR 876.5026
Third-party review
No
Summary or statement
Summary
Review time
215 daysreceipt to decision, calendar days
Median in QRC
228 daysthe latest 3 other decisions
QRC decisions
1in the last 12 months
Applicant’s 510(k)s
2in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The in2 Smart (in2S) is a wearable, non-implanted electrical stimulation device for premature ejaculation. It has reusable components that generate and control electrical stimulation, plus disposable Contact Pads that are applied to the perineum (the area between the genitals and anus) before intercourse.
What it’s cleared for
The in2 Smart is indicated for the management of premature ejaculation in males who ejaculate after intromission. It is designed to increase the time between arousal and ejaculation by delivering short, low-intensity electrical stimulation to the perineal muscles and nerves during intercourse. It is listed as an over-the-counter use.
Compared with (the predicate)
How it differs
The summary says the in2 Smart has a longer maximal stimulation duration (30 minutes, compared with 15 minutes for the vPATCH). It is sold as a kit with reusable parts and 10 single-use pads, and the user can personalize the intensity between Low and High, while the predicate offers two pre-set intensities.
Testing the summary lists
  • Hardware physical and functional testing, including basic electrical safety and EMC testing
  • Software verification and validation (IEC 62304:2006+A1:2015)
  • Usability testing
  • Biocompatibility testing (ISO 10993-1:2018; cytotoxicity, sensitization and irritation)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K252809 took 215 calendar days from receipt to decision. Of the latest 3 other decisions in product code QRC, 2 took longer, 1 was faster; their median was 228 days.

Latest decisions in QRC

NumberDevice and applicantDecision and review time
K252809 in2 Smart (in2S)Virility Medical , Ltd. · Traditional Apr 6, 2026215 days
K241897 MORMorari, Inc. · Traditional Feb 11, 2025228 days
K223595 vPATCHVirility Medical , Ltd. · Traditional May 1, 2023150 days
DEN210012 vPatchVirility Medical · Direct Nov 23, 2021238 days

Other 510(k)s from Virility Medical , Ltd.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K223595 vPATCHVirility Medical , Ltd. · Traditional May 1, 2023150 days

Watch QRC in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get QRC every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in QRC. Free. Add up to two more codes, separated by commas.