FDA device news · Product code MAX

FDA 510(k) clearance

UniSpace® TPLIF Cage (K252351): Cg Medtech Co., Ltd. 510(k) clearance

Uijeongbu-Si, KR

Follow MAX: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Oct 28, 2025
Received
Jul 29, 2025
510(k) type
Traditional
Product code
MAXIntervertebral Fusion Device With Bone Graft, Lumbar
Review panel
Orthopedic
Device class
Class II
Regulation
21 CFR 888.3080
Third-party review
No
Summary or statement
Summary
Review time
91 daysreceipt to decision, calendar days
Median in MAX
96 daysthe latest 24 other decisions
MAX decisions
21in the last 12 months
Applicant’s 510(k)s
2in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The UniSpace TPLIF Cage is a spinal implant used to fuse vertebrae in the lower back. It is a hollow cage made of titanium alloy (Ti6Al4V ELI) and built by 3D printing (direct metal laser sintering). It comes in several heights, widths, lengths and lordotic angles, which describes the built-in angle that restores the natural curve of the spine, and it has an open design that holds bone graft material.
What it’s cleared for
The device is cleared for intervertebral body fusion, which means joining two spinal bones, at one level or two neighboring levels in the lumbar spine from L2 to S1. It is for adult patients with degenerative disc disease and up to Grade I spondylolisthesis (slipping of one vertebra over another). The patients must be skeletally mature and have had six months of non-operative treatment. It must be used with supplemental internal spinal fixation and with autograft or allograft bone.
Compared with (the predicate)
  • K132926, MAX Velofix™ Interbody Fusion System
Testing the summary lists
  • Static compression test (ASTM F2077)
  • Static compression shear test (ASTM F2077)
  • Dynamic compression test (ASTM F2077)
  • Dynamic compression shear test (ASTM F2077)
  • Subsidence test (ASTM F2267)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K252351 took 91 calendar days from receipt to decision. Of the latest 24 other decisions in product code MAX, 14 took longer, 10 were faster; their median was 96 days.

Latest decisions in MAX

NumberDevice and applicantDecision and review time
K252865 4CIS BLUE DOLPHIN CAGE SYSTEMSolco Biomedical Company India Private Limited · Traditional Nov 6, 2026423 days
K262572 Symmetrix™ PX Spinal SystemMedtronic Sofamor Danek USA, Inc. · Traditional Oct 29, 202694 days
K262646 Calibrate MAX System; Calibrate NanoTec MAX System; Calibrate CCX Interbody System; Calibrate NanoTec CCX Interbody SystemAlphatec Spine, Inc. · Traditional Oct 21, 202683 days
K261049 SCALA® TLSpineart SA · Traditional Aug 20, 2026142 days
K261430 TruLift Expandable SystemLife Spine, Inc. · Traditional Jul 28, 202689 days
K253011 Catalyft PL Expandable Interbody SystemMedtronic Sofamor Danek USA, Inc. · Traditional Jul 17, 2026301 days
K261020 Symmetrix™ L Spinal SystemMedtronic Sofamor Danek USA, Inc. · Traditional Jul 14, 2026109 days
K253420 Admiral ACP System; Explorer TO System; Laminoplasty System; Manta Ray TDF Spacer; Meridian Interbody System; Meridian Anterior Plate System; Reef L Interbody System; Reef TO/TA System; Regatta Lateral System; Regatta Lateral Plate System; Shoreline ACS Interbody System; Shoreline Threaded TruProfile Plate; Shoreline RT Interbody System; Vu Mesh; WaveForm A Interbody System; WaveForm C Interbody System; WaveForm L Interbody System; WaveForm TA Interbody System; WaveForm TO Interbody SySeaSpine Orthopedics Corporation · Traditional Jul 1, 2026273 days
K253151 Fule Interbody Fusion Cage SystemBeijing Fule Science & Technology Development Co., Ltd. · Traditional Jun 15, 2026262 days
K261159 Exceed® Biplanar Expandable Interbody SystemSpine Wave, Inc. · Traditional May 29, 202651 days

Other 510(k)s from Cg Medtech Co., Ltd.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K251725 ANAX™ OCT Spinal SystemCg Medtech Co., Ltd. · Special Jun 25, 202520 days

Watch MAX in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get MAX every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in MAX. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.