FDA device news · Product code CHL

FDA 510(k) clearance

ABL90 FLEX PLUS System, safeCLINITUBES (K252207): Radiometer Medicals Aps 510(k) clearance

Brønshøj, DK

Follow CHL: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Apr 10, 2026
Received
Jul 14, 2025
510(k) type
Traditional
Product code
CHLElectrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Review panel
Clinical Chemistry
Device class
Class II
Regulation
21 CFR 862.1120
Third-party review
No
Summary or statement
Summary
Review time
270 daysreceipt to decision, calendar days
Median in CHL
249 daysthe latest 24 other decisions
CHL decisions
1in the last 12 months
Applicant’s 510(k)s
5in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The ABL90 FLEX PLUS System is a portable, automated blood gas analyzer made by Radiometer Medical ApS. It includes the analyzer, sensor cassette and solution pack consumables, and accessories. It uses potentiometry (an electrical measurement of voltage changes in a sample) to measure pH and pCO2, and spectrophotometry (measuring how much light a sample absorbs) to measure total hemoglobin (ctHb).
What it’s cleared for
The system is cleared to measure blood gas (pCO2) in arterial, venous and capillary whole blood, and pH and total hemoglobin (ctHb) in capillary whole blood. It is for use by trained technologists, nurses, physicians and therapists, in a laboratory, near patient or point-of-care setting, and only under a physician's order. The safeCLINITUBES are cleared to collect, mix and dispense capillary whole blood samples on the system.
Compared with (the predicate)
How it differs
The summary says the subject device is cleared for pCO2 in capillary, arterial and venous whole blood and for pH and ctHb in capillary whole blood. The predicate's listed indications also include electrolytes, glucose, lactate and neonatal bilirubin, which the subject device's indications do not list.
Testing the summary lists
  • Analytical performance per CLSI guidelines: linearity (CLSI EP06, EP39)
  • Detection capability, limit of quantitation (CLSI EP17-A2, EP39)
  • Precision using QC ampoules and whole blood (CLSI EP05-A3)
  • Method comparison and bias (CLSI EP09c)
  • Interference testing for Intralipid, hemolysis, bilirubin, biotin and propofol (CLSI EP07, EP37)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K252207 took 270 calendar days from receipt to decision. Of the latest 24 other decisions in product code CHL, 8 took longer, 16 were faster; their median was 249 days.

Latest decisions in CHL

NumberDevice and applicantDecision and review time
K252207 ABL90 FLEX PLUS System, safeCLINITUBESRadiometer Medicals Aps · Traditional Apr 10, 2026270 days
K244014 i-STAT CG4+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · Traditional May 2, 2025127 days
K240998 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · Traditional Dec 13, 2024246 days
K230285 i-STAT CG8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · Traditional Oct 27, 2023267 days
K221900 Stat Profile Prime Plus Analyzer SystemNova Biomedical Corporation · Traditional Sep 29, 2023456 days
K223857 i-STAT G3+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · Traditional Sep 15, 2023266 days
K223608 GEM Premier 7000 with IQM3Instrumentation Laboratory Company · Traditional Aug 10, 2023251 days
K211559 EasyStat 300Medica Corporation · Traditional Jan 27, 2023617 days
K203790 GEM Premier 5000Instrumentation Laboratory CO · Special Aug 20, 2021235 days
K200986 OPTI® B-Lac CassetteOpti Medical Systems, Inc. · Traditional Aug 2, 2021474 days

Other 510(k)s from Radiometer Medicals Aps

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K252475 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · Traditional May 1, 2026267 days
K252488 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · Traditional May 1, 2026266 days
K241037 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · Traditional Jan 14, 2025273 days
K240998 ABL90 FLEX PLUS SystemRadiometer Medicals Aps · Traditional Dec 13, 2024246 days

Watch CHL in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get CHL every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in CHL. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.