FDA 510(k) clearance
ROSA Knee System with UltraSound Imaging Platform (USIP) (K252058): Balmoral Medical, LLC 510(k) clearance
Rosemont, IL
- Decision
- Substantially Equivalent
- Decision date
- Feb 6, 2026
- Received
- Jul 1, 2025
- 510(k) type
- Traditional
- Product code
- OLOOrthopedic Stereotaxic Instrument
- Review panel
- Neurology
- Device class
- Class II
- Regulation
- 21 CFR 882.4560
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 220 daysreceipt to decision, calendar days
- Median in OLO
- 103 daysthe latest 25 other decisions
- OLO decisions
- 47in the last 12 months
- Applicant’s 510(k)s
- 1in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The ROSA Knee System with Ultrasound Imaging Platform (USIP) is a surgical stereotaxic instrument used to assist surgeons in total knee replacement (TKA). It includes a robotic arm, an optical tracking system, and an ultrasound-based add-on.
- What it’s cleared for
- The device is cleared as a stereotaxic instrumentation system to assist the surgeon in placing knee implant components during total knee replacement surgery in skeletally mature patients. Robotic arm placement is based on the tracked position of the femur and tibia and on the surgical plan set by the surgeon during the operation. It is used with specified fixed bearing Persona knee replacement systems. The device does not include a pre-operative planning feature.
- Compared with (the predicate)
- K182964, ROSA Knee System
- How it differs
- The summary says the main difference is how the position of the optical guide relative to the patient's bone is determined and tracked. The USIP uses continuous, non-invasive ultrasonic bone detection through soft tissue instead of pins placed in the bone. The summary also says the device uses B-mode ultrasound imaging, which the predicate does not include.
- Testing the summary lists
- Software verification and validation (IEC 62304; FDA software guidance)
- Biocompatibility (ISO 10993-1, 5th edition 2018-08)
- Reprocessing and cleaning validation (FDA reprocessing guidance)
- Electrical safety and EMC (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37)
- Acoustic output and thermal/mechanical indices (IEC 62359)
- Accuracy verification and cadaveric validation of imaging performance
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K252058 took 220 calendar days from receipt to decision. Of the latest 25 other decisions in product code OLO, 2 took longer, 23 were faster; their median was 103 days.
Latest decisions in OLO
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262567 | GPS AnkleBlue Ortho · Traditional | Oct 26, 202694 days |
| K260403 | 7D Surgical System (10-0001)7D Surgical ULC · Traditional | Oct 19, 2026255 days |
| K263234 | TMINI Miniature Robotic SystemTHINK Surgical, Inc. · Special | Oct 16, 202630 days |
| K263154 | CarboClear® Cervical Navigated InstrumentsCarboFix Orhtopedics , Ltd. · Special | Oct 9, 202630 days |
| K262257 | Spine Planning (3.0); Spine & Trauma Navigation; Alignment System SpineBrainlab SE · Traditional | Sep 30, 202690 days |
| K262506 | Remex Spine Surgery Navigation - Universal InstrumentsRemex Medical Corp. · Special | Aug 20, 202630 days |
| K261410 | Navbit MARQ (PRD-001-SUP (Supine Registration Kit), PRD-001-LAT (Lateral Registration Kit))Suite 201, National Innovation Centre · Traditional | Jul 30, 202691 days |
| K261205 | SPINEART Navigation Instrument SystemSpineart SA · Traditional | Jul 29, 2026107 days |
| K261024 | TMINI Miniature Robotic SystemTHINK Surgical, Inc. · Traditional | Jul 24, 2026116 days |
| K262121 | Ion-L Navigated InstrumentsSurGenTec, LLC · Special | Jul 24, 202631 days |
Other 510(k)s from Balmoral Medical, LLC
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
This is the only 510(k) in openFDA under this applicant name.
Watch OLO in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get OLO every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in OLO. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K252058Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:OLORetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:OLORetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:OLO AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Balmoral Medical, LLC"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil