FDA device news · Product code FAS

FDA 510(k) clearance

Dornier Bi-Polar Electrode BIP12CLM Bipolar 12° Medium Cutting Loop 24 Fr Electrode-Sterile, Single-Use BIP12CLL Bipolar 12° Large Cutting Loop 24 Fr Electrode-Sterile, Single-Use BIP30CLM Bipolar 30° Medium Cutting Loop 24 Fr Electrode-Sterile, Single-Use BIP30CLL Bipolar 30° Large Cutting Loop 24 Fr Electrode-Sterile, Single-Use BIPMBLA Bipolar Medium Bladder Loop 24 Fr Electrode-Sterile, Single-Use BIPNEEL Bipolar Needle 24 Fr Electrode-Sterile, Single-Use BIPDDSC Bipolar Disc ) (K251403): Dornier Medtech America Inc (Dmta) 510(k) clearance

Kennesaw, GA

Follow FAS: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Dec 5, 2025
Received
May 6, 2025
510(k) type
Traditional
Product code
FASElectrode, Electrosurgical, Active, Urological
Review panel
Gastroenterology, Urology
Device class
Class II
Regulation
21 CFR 876.4300
Third-party review
No
Summary or statement
Summary
Review time
213 daysreceipt to decision, calendar days
Median in FAS
102 daysthe latest 24 other decisions
FAS decisions
1in the last 12 months
Applicant’s 510(k)s
1in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Dornier Bi-Polar Electrode is a sterile, single-use accessory that attaches to a compatible resectoscope, a thin telescope-like instrument inserted through the urethra. It connects to a high-frequency electrosurgical generator and delivers electrical energy at the treatment site, using a plasma field in saline to cut, coagulate, and remove soft tissue.
What it’s cleared for
The device is cleared for endoscopic urological procedures in the prostate, bladder, and bladder neck. It is used for resection, ablation, or removal of soft tissue where hemostasis is required, including procedures such as TURP, TUIP, and TURBT, and for vaporization procedures. It is intended for use in an irrigated environment and is not intended for treating prostate cancer.
Compared with (the predicate)
  • K151976, Omnitech HF Resection/Vaporization Electrode
How it differs
The summary says the indications for use are the same as the predicate. The only significant technological difference it identifies is the use of tungsten, rather than platinum/iridium, to form the loop and needle cutting surfaces. Both devices are single use and sterilized with ethylene oxide.
Testing the summary lists
  • Bench performance testing: dimensional analysis, mechanical tensile strength, thermal penetration, and simulated use
  • Electrical safety (IEC 60601-1, IEC 60601-1-2:2014)
  • Biocompatibility (ISO 10993-5, -23, -10, -11)
  • Sterilization validation (EN ISO 11135:2014, ethylene oxide)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K251403 took 213 calendar days from receipt to decision. Of the latest 24 other decisions in product code FAS, 2 took longer, 22 were faster; their median was 102 days.

Latest decisions in FAS

NumberDevice and applicantDecision and review time
K251403 Dornier Bi-Polar Electrode BIP12CLM Bipolar 12° Medium Cutting Loop 24 Fr Electrode-Sterile, Single-Use BIP12CLL Bipolar 12° Large Cutting Loop 24 Fr Electrode-Sterile, Single-Use BIP30CLM Bipolar 30° Medium Cutting Loop 24 Fr Electrode-Sterile, Single-Use BIP30CLL Bipolar 30° Large Cutting Loop 24 Fr Electrode-Sterile, Single-Use BIPMBLA Bipolar Medium Bladder Loop 24 Fr Electrode-Sterile, Single-Use BIPNEEL Bipolar Needle 24 Fr Electrode-Sterile, Single-Use BIPDDSC Bipolar Disc )Dornier Medtech America Inc (Dmta) · Traditional Dec 5, 2025213 days
K250522 Multi4 SystemMulti4 Medical AB · Traditional Jun 27, 2025126 days
K240504 PLASMA EDGE SystemLamidey Noury Medical · Special Jun 12, 2024112 days
K213135 Plasma Edge resection and vaporization electrodes, Plasma Edge working elements, Adaptor Plasma EdgeChirurgie Innovation · Special Mar 2, 2022156 days
K210651 Resection Electrodes with HF cableOlympus Winter & Ibe GmbH · Traditional Aug 3, 2021152 days
K210394 Electrosurgical accessoryDornier Medtech America, Inc. · Traditional May 19, 202198 days
K171965 Resection Electrodes with HF cableOlympus Winter & Ibe GmbH · Traditional Sep 28, 201790 days
K163090 Plasma Edge SystemChirurgie Innovation · Traditional May 23, 2017200 days
K151976 Omnitech HF Resection Electrodes, Omnitech HF Vaporization ElectrodeOmnitech Systems, Inc. · Traditional Jun 27, 2016347 days
K152092 Resection ElectrodesOlympus Winter & Ibe GmbH · Traditional Oct 26, 201590 days

Other 510(k)s from Dornier Medtech America Inc (Dmta)

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

This is the only 510(k) in openFDA under this applicant name.

Watch FAS in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get FAS every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in FAS. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.