FDA device news · Product code OWB
FDA 510(k) clearance
Allia Moveo (K251199): GE Medical Systems SCS 510(k) clearance
Buc, FR
- Decision
- Substantially Equivalent
- Decision date
- Dec 9, 2025
- Received
- Apr 18, 2025
- 510(k) type
- Traditional
- Product code
- OWBInterventional Fluoroscopic X-Ray System
- Review panel
- Radiology
- Device class
- Class II
- Regulation
- 21 CFR 892.1650
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 235 daysreceipt to decision, calendar days
- Median in OWB
- 177 daysthe latest 24 other decisions
- OWB decisions
- 18in the last 12 months
- Applicant’s 510(k)s
- 40in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Allia Moveo is an interventional X-ray system made by GE HealthCare. It has a C-arm, a movable arm that holds the X-ray tube on one side and a digital detector on the other, mounted on a motorized cart, along with an X-ray table. It produces fluoroscopic images, which are live X-ray moving pictures, and rotational images, which are X-ray views taken as the C-arm turns around the patient.
- What it’s cleared for
- The device is cleared to generate fluoroscopic and rotational X-ray images of human anatomy for cardiovascular, vascular and non-vascular diagnostic and interventional procedures, for patients from newborn to geriatric. With the OR table, it is also cleared to generate such images for image-guided surgical procedures, and the OR table is suitable for interventional and surgical procedures.
- Compared with (the predicate)
- K232344, Allia IGS 7
- How it differs
- The summary says the new device introduces a new C-arm with a modified detector mount. It uses a 41x41 cm detector, while the predicate uses a 31x31 cm detector.
- Testing the summary lists
- Electrical safety and standards conformance (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, IEC 60601-2-46, NEMA XR-27)
- Engineering bench testing of the new gantry using anthropomorphic phantoms
- Cybersecurity and software documentation under FDA guidance
- Pediatric information for X-ray imaging device premarket notifications under FDA guidance
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K251199 took 235 calendar days from receipt to decision. Of the latest 24 other decisions in product code OWB, 7 took longer, 17 were faster; their median was 177 days.
Latest decisions in OWB
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253726 | uAngio AVIVA CXShanghai United Imaging Healthcare Co., Ltd. · Traditional | Aug 21, 2026270 days |
| K261413 | EXTRON 3; EXTRON 6DRTECH Corporation · Traditional | Aug 19, 2026111 days |
| K254205 | Zenition 30Philips Medical Systems Nederland B.V. · Traditional | Aug 7, 2026221 days |
| K260607 | Shield System (Siemens 30x40)Radiaction , Ltd. · Special | Jul 31, 2026157 days |
| K253670 | SmartCT (R3.0 (L10))Philips Medical Systems B.V. · Abbreviated | Jul 10, 2026231 days |
| K252735 | C-beamPulmera, Inc. · Traditional | May 22, 2026267 days |
| K252503 | Intelligent NRCanon, Inc. · Traditional | Apr 30, 2026265 days |
| K254173 | ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoSiemens Medical Solutions USA, Inc. · Traditional | Apr 24, 2026122 days |
| K253752 | ARTIS genio floor; ARTIS icono.explore floorSiemens Medical Solutions USA, Inc. · Traditional | Apr 23, 2026149 days |
| K252229 | BELLIGER ACEGenoray Co., Ltd. · Traditional | Apr 2, 2026260 days |
Other 510(k)s from GE Medical Systems SCS
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253770 | CardIQ SuiteGE Medical Systems SCS · Traditional | Jul 21, 2026237 days |
| K260087 | Senographe Pristina; SenoBright HD; Pristina Serena; Pristina Serena BrightGE Medical Systems SCS · Traditional | Mar 24, 202671 days |
| K243672 | CardIQ SuiteGE Medical Systems SCS · Traditional | Jun 18, 2025203 days |
| K243651 | VersaViewerGE Medical Systems SCS · Traditional | Apr 21, 2025146 days |
| K243446 | 3DXRGE Medical Systems SCS · Traditional | Feb 25, 2025111 days |
| K233731 | CardIQ SuiteGE Medical Systems SCS · Traditional | Aug 1, 2024254 days |
| K233714 | BreViewGE Medical Systems SCS · Traditional | Jan 25, 202466 days |
| K232344 | Allia IGS 3, Allia IGS 5, Allia IGS 7, Allia IGS 7 ORGE Medical Systems SCS · Special | Oct 11, 202368 days |
| K223424 | Spine Auto ViewsGE Medical Systems SCS · Traditional | Jul 13, 2023241 days |
| K223490 | FlightPlan for EmbolizationGE Medical Systems SCS · Traditional | Mar 21, 2023120 days |
| K222895 | DynamicIQGE Medical Systems SCS · Traditional | Nov 22, 202260 days |
| K223152 | Vision 2, EVARVision, TrackVision 2, HeartVision 2GE Medical Systems SCS · Traditional | Nov 22, 202247 days |
Watch OWB in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get OWB every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in OWB. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K251199Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:OWBRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:OWBRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:OWB AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"GE Medical Systems SCS"Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil