FDA device news · Product code MEH

FDA 510(k) clearance

Allure Hip Stem and Intramedullary Plugs (K250375): Waldemar Link GmbH & Co. KG 510(k) clearance

Hamburg, DE

Follow MEH: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Nov 5, 2025
Received
Feb 10, 2025
510(k) type
Traditional
Product code
MEHProsthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Review panel
Orthopedic
Device class
Class II
Regulation
21 CFR 888.3353
Third-party review
No
Summary or statement
Summary
Review time
268 daysreceipt to decision, calendar days
Median in MEH
108 daysthe latest 24 other decisions
MEH decisions
1in the last 12 months
Applicant’s 510(k)s
43in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Allure Hip Stem and Intramedullary Plugs are orthopedic implants for hip replacement. The hip stem is a metal femoral component (titanium alloy, Ti6Al4V) with a coating meant to help it attach to bone without cement. The Intramedullary Plugs are small polyethylene plugs (UHMWPE, a tough, wear-resistant plastic) placed in the thigh bone canal to keep bone cement from spreading where it should not go.
What it’s cleared for
The Allure Hip Stems are cleared for uncemented total and partial (hemi) hip replacement to treat mobility-limiting hip or upper thigh bone problems that cannot be treated with non-surgical or bone-fixation methods. Listed indications include osteoarthritis, rheumatoid arthritis, correction of functional deformities, avascular necrosis, femoral neck fractures and revision surgery.
Compared with (the predicate)
  • K150862, DePuy Orthopaedics, Actis DuoFix Hip Prosthesis
  • K160907, DePuy Orthopaedics, Actis DuoFix Hip Prosthesis
  • K210581, DePuy Orthopaedics, Actis DuoFix Hip Prosthesis
  • K211703, Stryker Orthopaedics, Insignia Hip Stem
How it differs
The summary says the Allure Stem's proximal coating is plasma-sprayed titanium, while the primary predicate (Actis DuoFix) uses sintered titanium beads. The summary notes that plasma spray is a common coating process and is also used on the additional predicate stem.
Testing the summary lists
  • Fatigue testing of the stem (ISO 7206-4 and ISO 7206-6)
  • Range of motion evaluation (ISO 21535)
  • Engineering analyses of fretting corrosion and impingement
  • Coating characterization and Intramedullary Plug verification testing
  • Biocompatibility (ISO 10993) and sterilization packaging simulation (ASTM D4169)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K250375 took 268 calendar days from receipt to decision. Of the latest 24 other decisions in product code MEH, 3 took longer, 21 were faster; their median was 108 days.

Latest decisions in MEH

NumberDevice and applicantDecision and review time
K261576 Icona Hip StemCorin, Ltd. · Traditional Aug 10, 202690 days
K250375 Allure Hip Stem and Intramedullary PlugsWaldemar Link GmbH & Co. KG · Traditional Nov 5, 2025268 days
K251166 Mfinity Femoral SystemMedacta International S.A. · Traditional Jun 13, 202559 days
K251052 Trivicta® Hip StemOrtho Development Corp. · Traditional May 22, 202549 days
K242315 Resolve Modular Revision Hip StemUnited Orthopedic Corporation · Traditional May 1, 2025269 days
K243839 Alteon® HA Femoral StemsExactech, Inc. · Traditional Mar 10, 202587 days
K243162 World LinerSignature Orthopaedics Pty, Ltd. · Special Nov 25, 202456 days
K241472 Icona Hip StemCorin U.S.A. Limited · Traditional Oct 3, 2024132 days
K241875 TheRay Collared and Collarless Femoral StemNextstep Arthropedix · Traditional Sep 13, 202477 days
K241690 Logical Liner; World Liner; World Knee PatellaSignature Orthopaedics Pty, Ltd. · Special Aug 1, 202450 days

Other 510(k)s from Waldemar Link GmbH & Co. KG

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K261009 MobileLink Acetabular Cup System – 40mm ID Inserts (various)Waldemar Link GmbH & Co. KG · Traditional Jul 6, 2026102 days
K243927 MobileLink Acetabular Cup System - inhouse coatingsWaldemar Link GmbH & Co. KG · Traditional May 16, 2025147 days
K241636 MobileLink Acetabular Cup System - Line Extension (Multiple)Waldemar Link GmbH & Co. KG · Traditional Nov 8, 2024155 days
K231445 LINK Embrace Shoulder System - Reverse ConfigurationWaldemar Link GmbH & Co. KG · Traditional Nov 16, 2023182 days
K222066 LINK MobileLink Acetabular Cup SystemWaldemar Link GmbH & Co. KG · Traditional Apr 14, 2023275 days
K230471 LinkSymphoKnee – Fixed Bearing Ultracongruent (FB UC) Articulating SurfaceWaldemar Link GmbH & Co. KG · Special Mar 21, 202328 days
K213770 SP-CL Hip Stem and LCU Hip SystemWaldemar Link GmbH & Co. KG · Traditional Sep 13, 2022285 days
K221794 Vario-Cup SystemWaldemar Link GmbH & Co. KG · Special Jul 20, 202229 days
K211567 BiMobile Instruments (for BiMobile Dual Mobility System)Waldemar Link GmbH & Co. KG · Special Jun 3, 2022378 days
K220628 LINK Endo-Model EVO Knee SystemWaldemar Link GmbH & Co. KG · Traditional May 3, 202260 days
K213675 MP Reconstruction SystemWaldemar Link GmbH & Co. KG · Special Apr 11, 2022140 days
K212992 LINK® Embrace Shoulder System - Reverse ConfigurationWaldemar Link GmbH & Co. KG · Traditional Nov 18, 202159 days

Watch MEH in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get MEH every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in MEH. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.