FDA device news · Product code MEB

FDA 510(k) clearance

SMARTeZ™ Elastomeric Infusion Pump (RS series) (K250234): Epic Medical Pte. , Ltd. 510(k) clearance

Singapore, SG

Follow MEB: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Apr 24, 2026
Received
Jan 27, 2025
510(k) type
Traditional
Product code
MEBPump, Infusion, Elastomeric
Review panel
General Hospital
Device class
Class II
Regulation
21 CFR 880.5725
Third-party review
No
Summary or statement
Summary
Review time
452 daysreceipt to decision, calendar days
Median in MEB
88 daysthe latest 24 other decisions
MEB decisions
1in the last 12 months
Applicant’s 510(k)s
22in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The SMARTeZ Elastomeric Infusion Pump (RS series) is a sterile, single-use, non-electronic pump for continuous drug delivery. It uses a stretchy elastomeric (rubber-like) reservoir that, when filled, pushes its own elastic recoil to drive fluid through the tubing and a flow restrictor to the patient.
What it’s cleared for
The device is cleared for continuous infusion of medications. The Long infusion time article is for general infusion use, including pain management, by intravenous or subcutaneous routes. The Short infusion time article is for general infusion use, including antibiotic delivery, by intravenous route. The Chemotherapy article is for continuous infusion of chemotherapy medications by intravenous or intra-arterial routes. The summary states the device is for prescription use.
Compared with (the predicate)
  • K151650, SMARTeZ Elastomeric Infusion Pump
Testing the summary lists
  • Functional performance to ISO 28620:2020 and ISO 80369-7:2021 luer connector tests, plus ANSI/AAMI CN27:2021 for luer activated valves
  • Sterilization (ethylene oxide) to ISO 11135:2014
  • Shipping and package integrity (ASTM D 4169-16, ASTM F1980-21, ASTM F88/F88M-21, ASTM F1929-23, EN 868-5:2009)
  • Pyrogen testing (ANSI/AAMI ST72/2019, USP <151>, <161>, <85>), performed on every lot
  • Shelf-life of 3 years (36 months) validated using accelerated aging

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K250234 took 452 calendar days from receipt to decision. Of the latest 24 other decisions in product code MEB, 24 were faster; their median was 88 days.

Latest decisions in MEB

NumberDevice and applicantDecision and review time
K250234 SMARTeZ™ Elastomeric Infusion Pump (RS series)Epic Medical Pte. , Ltd. · Traditional Apr 24, 2026452 days
K240624 InfuLifeFirst Medical Source, LLC · Traditional Nov 5, 2024245 days
K242152 SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte. , Ltd. · Special Oct 7, 202476 days
K181360 ON-Q* Pump with BolusHalyard Health · Traditional Mar 22, 2019304 days
K151650 SMARTez elastomeric infusion pumpEpic Medical Pte. , Ltd. · Traditional Apr 4, 2016291 days
K131249 ON-Q PAINBUSTER, ON-Q C-BLOC, HOMEPUMP ECLIPSE, HOMEPUMP C-SERIESI-Flow, LLC · Traditional Feb 3, 2014278 days
K131247 MULTIBOLUS II DISPOSABLE PAIN MANAGEMENT SYSTEMMfs - Medical Flow Systems, Ltd. · Traditional Aug 14, 2013105 days
K090300 MODIFICATION TO AUTOFUSERAce Medical Us, LLC · Special Apr 30, 200983 days
K081905 ACCUFLOW AND ACCUFLUX ELASTOMERIC INFUSION PUMPWestmed, Inc. · Traditional Jul 15, 200812 days
K071532 ALPHA INFUSION PUMP, MODEL A200, A450Advanced Infusion, Inc. · Traditional Feb 19, 2008259 days

Other 510(k)s from Epic Medical Pte. , Ltd.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K252987 ADRx™ Admixture DeviceEpic Medical Pte. , Ltd. · Traditional Jun 12, 2026267 days
K251722 ProSeal™ Transfer Injector (421120, 421130, 421140)Epic Medical Pte. , Ltd. · Special Dec 5, 2025184 days
K251814 EZ™ IV Administration SetEpic Medical Pte. , Ltd. · Traditional Aug 29, 202577 days
K252094 eZSURE™ Empty Fluid ContainerEpic Medical Pte. , Ltd. · Special Aug 1, 202529 days
K251340 ProSeal™ Bag Spike with Additive Port (423370ST, 423370)Epic Medical Pte. , Ltd. · Special May 29, 202529 days
K243976 ZeroClear™ Bag Access (423100)Epic Medical Pte. , Ltd. · Special Jan 22, 202530 days
K240517 ProSeal™ Vented Universal Vial AdaptorEpic Medical Pte. , Ltd. · Traditional Dec 26, 2024307 days
K241988 ProSeal™ Closed System Bag AccessEpic Medical Pte. , Ltd. · Traditional Oct 23, 2024107 days
K241735 ProSeal™ In Line Pump Set (423850)Epic Medical Pte. , Ltd. · Special Oct 10, 2024115 days
K242152 SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte. , Ltd. · Special Oct 7, 202476 days
K241823 ToxiSeal™ Vial AdaptorEpic Medical Pte. , Ltd. · Special Aug 30, 202467 days
K241442 eZSURE™ Empty Fluid Container with ProSeal™ Injection SiteEpic Medical Pte. , Ltd. · Special Jun 21, 202430 days

Watch MEB in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get MEB every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in MEB. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.